Mass spectrometers : puriFlash® MS
Mass spectrometers : puriFlash® MS
Configure your
puriFlash® MS
puriFlash® MS-HWM
& MS Splitter
Basic configuration
- Mass-triggered faction collection via normal or reverse-phase Flash or preparative HPLC
- Confirmation of the mass and purity of the collected fractions
- Real-time reaction monitoring
- Compact mass spectrometer that can be positioned underneath a fume hood, direct analyses.
- Source ESI, APCI or APCI/ASAP (Option)
- A single analysis for positive and negative ion mode
- Flow ESI: 10 µL/min to 1 mL/min
- Flow APCI: 10 µL/min to 2 mL/min
- Flow ASAP: 10 µL/min to 2 mL/min
- puriFlash® MS range: 10 to 1200 m/z
- puriFlash® MS-HMW range: 10 to 2000 m/z
- Acquisition frequency per second 10,000 m/z
- Sensitivity (ESI) 10 pg reserpin (FIA – injection of 5 µL to 100 µL/min)
- S/N: 100/1 (RMS) with SIM (Single Ion Monitoring) of 609.3 m/z
- Accuracy +/- 0.1 m/z for the entire acquisition range
- Stability +/- 0.1 m/z for a period of 12h (18°C -24°C)
- Polarity-change speed: 50 ms
- Direct injection (FIA) or ASAP source
- MS Splitter:
- ● Interchim® concept
- ● Combined “Split and Dilution” Technology
- ● Compatible with all column sizes
- ● Even with a high flow, no counterpressure is generated
- ● Dilution after split to adjust the quantity entering the source and to avoid system saturation
- Mass: connection to puriFlash® via USB
- MS Splitter: connection to puriFlash® via RS-232 cable
- Voltage to ground: 100/240 V AC voltage
- Outlet of choice: UK, Europe, US